Loading MedWatcher information…
Loading MedWatcher information…
FDA · FDA event 96085 · Posted Feb 4, 2025
Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-1019-2025 | Z-1019-2025 | UDI-DI: 00811505030122. Software 5.9.2 and prior | Not provided in the normalized record | Ongoing |
Jan 10, 2025
Recall initiated
Date recorded by the FDA.
Feb 4, 2025
Posted by FDA
The recall event appeared in the FDA source.
Oct 8, 2026
Last checked
MedWatcher last reconciled this record with the openFDA export.