Loading MedWatcher information…
Loading MedWatcher information…
FDA · FDA event 99589 · FDA posted Sep 29, 2026
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-3330-2026 | Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No | Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only) | Not provided in the normalized record | Ongoing |
Jul 16, 2026
Recall initiated
Date recorded by the FDA.
Sep 29, 2026
Posted by FDA
The recall event appeared in the FDA source.
Oct 28, 2026
Last checked
MedWatcher last reconciled this record with the openFDA export.