FDA · FDA event 98888 · Posted Jun 23, 2026
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-2525-2026 | Z-2525-2026 | REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613 | Not provided in the normalized record | Ongoing |
May 8, 2026
Recall initiated
Date recorded by the FDA.
Jun 23, 2026
Posted by FDA
The recall event appeared in the FDA source.