ICU Medical Plum 360 Infusion System
ICU Medical, Inc. · Infusion pumps
Active recall recordLast checked Oct 8, 2026
Loading MedWatcher information…
MedWatcher tracks the reviewed device family ICU Medical Plum 360 Infusion System under the canonical manufacturer identity ICU Medical, Inc.. The recall records on this page are limited to 7 owner-approved FDA recall actions.
This page covers only the reviewed device families and FDA recall actions shown below. It is not a complete product catalogue, and a recall does not necessarily affect every unit or configuration in a device family.
The manufacturer and device-family relationships come from the reviewed canonical-device inventory. Recall coverage is limited to the exact candidate IDs approved and published in the FUS-265 cohort; complete regulator payloads remain in the private source ledger.
Source databases can use a legal name, brand, division, or shortened name. MedWatcher retains the form used as evidence rather than treating every similar name as the same manufacturer.
Page reviewed Aug 17, 2026 by fusiondesigner@gmail.com · Next review by Nov 17, 2026.
ICU Medical, Inc. · Infusion pumps
Active recall recordLast checked Oct 8, 2026
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the…
ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback…
ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.
ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have…
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions.