ICU Medical Plum 360 Infusion System
ICU Medical, Inc. · Infusion pumps
Active recall recordLatest source check Oct 27, 2026
Loading MedWatcher information…
Devices and FDA recall records listed on MedWatcher for ICU Medical, Inc..
This page covers the devices and recall records listed below, not the complete product catalogue or recall history. A recall may apply only to specific models, lots, or serial numbers.
Device and manufacturer names are linked using official product, UDI, or FDA records. The source links below show the supporting information.
Official records may use a legal name, brand, division, or shortened name. Similar names do not necessarily refer to the same manufacturer.
ICU Medical, Inc. · Infusion pumps
Active recall recordLatest source check Oct 27, 2026
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the…
ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback…
ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.
ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have…
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions.