Loading MedWatcher information…
Loading MedWatcher information…
FDA · FDA event 98128 · Posted Jan 16, 2026
Update delayed: We have not successfully checked this official source since 2026-10-08 18:58:41 UTC. This page may not include the latest recall or safety information. Check the linked FDA source before making a health decision.
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-1120-2026 | Z-1120-2026 | 400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0 | Not provided in the normalized record | Ongoing |
| Z-1119-2026 | Z-1119-2026 | 400020403 (US), Software Versions: 1.2.2; 400021003 (Philippines), Software Versions: 1.2.0 | Not provided in the normalized record | Ongoing |
| Z-1118-2026 | Z-1118-2026 | 400020401 (US), Software Versions: 1.1.1, 1.1.2, 1.1.3, 1.1.4 400021001 (Philippines), Software Versions: 1.1.0 | Not provided in the normalized record | Ongoing |
Dec 15, 2025
Recall initiated
Date recorded by the FDA.
Jan 16, 2026
Posted by FDA
The recall event appeared in the FDA source.
Oct 8, 2026
Last checked
MedWatcher last reconciled this record with the openFDA export.