FDA · FDA event 97067 · Posted Aug 6, 2025
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-2180-2025 | Z-2180-2025 | REF: 10013186/ UDI: 07613203021173; REF: 2452-0007/ UDI: 10885403219870; REF: 2204-0007/ UDI: 10885403199363; REF: 24001-0007/ UDI: 10885403238666; REF: 10015012/ UDI: 10885403233951; REF: 2432-0007/ UDI: 10885403232329; REF: 2434-0007/ UDI: 07613203019682; REF: 2202-0007/ UDI: 10885403274039; REF: 11426965/ UDI: 10885403232558; REF: 10561554/ UDI: 10885403232565; REF: 10863358/ UDI: 10885403232466; REF: 11171447/ UDI: 07613203021234; REF: 11532269/ UDI: 10885403232343; REF: 10010454/ UDI: 07613203015806; REF: 2465-0007/ UDI: 10885403221941; REF: 10010453/ UDI: 07613203021135 REF: 11419365/ UDI: 10885403234736 REF: 10813621/UDI: 10885403234514 REF: 2423-0007/UDI: 07613203020985 REF: 10015896/UDI: 10885403234064 REF: 24010-0007T/UDI: 10885403271021 REF: 10010483/UDI: 10885403232602 REF: 2403-0007/UDI: 10885403232473 REF: 24301-0007T/UDI: 10885403223198 | Not provided in the normalized record | Ongoing |
| Z-2179-2025 | Z-2179-2025 | BD Alaris" Pump Module Model 8100- All Serial Numbers/ UDI(s): 10885403222054, 10885403517723, 10885403810015, 10885403810039, 10885403810046 | Not provided in the normalized record | Ongoing |
Jul 8, 2025
Recall initiated
Date recorded by the FDA.
Aug 6, 2025
Posted by FDA
The recall event appeared in the FDA source.
Oct 8, 2026
Last checked
MedWatcher last reconciled this record with the openFDA export.