FDA · FDA event 97669 · Posted Nov 20, 2025
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-0430-2026 | Z-0430-2026 | All Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743 | Not provided in the normalized record | Ongoing |
Oct 17, 2025
Recall initiated
Date recorded by the FDA.
Nov 20, 2025
Posted by FDA
The recall event appeared in the FDA source.
Oct 8, 2026
Last checked
MedWatcher last reconciled this record with the openFDA export.