FDA · FDA event 96945 · Posted Jun 24, 2025
Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-2016-2025 · all | Z-2016-2025 | Model Number: PT-000664/M009-S-AP. UDI-DI: 10385083000336. All software versions 1.X.X | Not provided in the normalized record | Ongoing |
May 23, 2025
Recall initiated
Date recorded by the FDA.
Jun 24, 2025
Posted by FDA
The recall event appeared in the FDA source.
Oct 8, 2026
Last checked
MedWatcher last reconciled this record with the openFDA export.