FDA · FDA event 97597 · Posted Oct 10, 2025
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-0132-2026 | Z-0132-2026 | Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X | Not provided in the normalized record | Ongoing |
Sep 11, 2025
Recall initiated
Date recorded by the FDA.
Oct 10, 2025
Posted by FDA
The recall event appeared in the FDA source.
Oct 8, 2026
Last checked
MedWatcher last reconciled this record with the openFDA export.