Dexcom G6 Continuous Glucose Monitoring System
Dexcom, Inc. · Continuous glucose monitors
Active recall recordLatest source check Oct 27, 2026
Loading MedWatcher information…
Devices and FDA recall records listed on MedWatcher for Dexcom, Inc..
Visit official website (opens in a new tab)This page covers the devices and recall records listed below, not the complete product catalogue or recall history. A recall may apply only to specific models, lots, or serial numbers.
Device and manufacturer names are linked using official product, UDI, or FDA records. The source links below show the supporting information.
Official records may use a legal name, brand, division, or shortened name. Similar names do not necessarily refer to the same manufacturer.
Dexcom, Inc. · Continuous glucose monitors
Active recall recordLatest source check Oct 27, 2026
Dexcom, Inc. · Continuous glucose monitors
Active recall recordLatest source check Oct 27, 2026
Dexcom, Inc. · Continuous glucose monitors
No linked recall recordsLatest source check Oct 27, 2026
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for…
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay"…
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a…
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier),…
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to…
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts…
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be…