FDA · FDA event 96743 · Posted Jun 16, 2025
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.
The official notice takes precedence over this summary. Talk to your clinician before changing prescribed therapy.
| Model | Description | Lot / serial range | Manufactured | Status |
|---|---|---|---|---|
| Z-1946-2025 | Z-1946-2025 | All Lots/ UDI: STK-DO-006 -00386270003188 STK-DO-013 -00386270002327 STK-DO-103-00386270002969 STK-DO-109 -00386270002792 | Not provided in the normalized record | Ongoing |
| Z-1945-2025 | Z-1945-2025 | All Lots/ UDI: STK-D7-013 - 00386270003812; STK-D7-014 - 00386270003843; STK-D7-103 - 00386270003836 | Not provided in the normalized record | Ongoing |
| Z-1944-2025 | Z-1944-2025 | All Lots/ UDI: STK-AT-011 - 00386270002198 STK-AT-012 -00386270002518 STK-AT-013 - 00386270002525 STK-GT-001 - 00386270004291 STK-GT-008 - 00386270004260 STK-GT-013 - 00386270002273 STK-GT-019 - 04595986293080 STK-GT-100 - 00386270002020 STK-GT-109 - 00386270002129 STK-GT-113 - 00386270003638 | Not provided in the normalized record | Ongoing |