Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, recall — FDA event 97971
The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.
Loading MedWatcher information…
Records aggregated from the official sources MedWatcher monitors. Each record links back to its source with the agency identifier and dates. Sample records are labeled.
109 recall records
The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.
Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.
Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested,…
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer…
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a…